Basic Operation

67

After the line disappears, the trigger level is marked with a

small left arrow.

Auto and Normal Trigger Modes

Press the trigger MENU key to display the TRIGGER menu

and press the Mode softkey to select Auto or Normal trigger

mode.

Auto mode

Use the auto trigger mode for signals other than

low-repetitive-rate signals and for unknown signal levels. To

display a DC signal, you must use Auto trigger mode since

there is no edge to trigger on.

When you press RUN/STOP key to start acquiring, the

oscilloscope first fill the pre-trigger buffer. It starts to search

for a trigger after the pre-trigger buffer is filled, and continues

to flow data through this buffer while it searches for the trigger.

While searching for the trigger, the oscilloscope overflows the

pre-trigger buffer; the first data put into the buffer is the first

pushed out. When a trigger is found, the pre-trigger buffer will

contain the events that occurred just before the trigger. If no

trigger is found, the oscilloscope generates a trigger and

displays the data as though a trigger had occurred. In this

case, the background of the Auto indicator at the top of the