Power requirements, Display annunciators – B&K Precision 8500-8526 - Manual User Manual

Page 11

Advertising
background image

not covered by warranty.

4

Line voltage selection switch (110 VAC or *220 VAC)
*For models 8524 and 8526, the line voltage switch is not usable and does not
allow for 220VAC operation. Do not connect them directly into a 220V line. For
220V operation, please order model 8524-220V or 8526-220V.

Power Requirements

All of the models, except for models 8524 and 8526, can accept 110V or 220V line input. To switch
between the two line voltage inputs:

1.

Verify and change the fuse inside the fuse box beneath the AC input receptacle. See fuse
section for reference to the fuse type and rating.

2.

Flip the 110V/220V line selection switch in the rear panel of the instrument.

WARNING:
Models 8524 and 8526 are configured for 110 V operation only. These two models are not
configurable via the line switch on the rear panel. Connecting an AC line input of a 8524 and
8526 to a 220V AC source will damage the units. If 220 V operation is desired, order model 8524-
220V or 8526-220V. “-220V” denotes 220 V operation only, reconfiguration to 110 V is not
possible.

Display annunciators

Annunciator

Meaning

OFF

The load is off.

UNREG

The input is unregulated.

CC

Constant current mode.

CV

Constant voltage mode.

CW

Constant power mode.

CR

Constant resistance mode.

PROT

Not used.

TRAN

Transient operation is enabled.

LIST

List mode is initiated or running.

SENSE

Remote sensing is on.

8500 DC Load Series

Version: 030614

Page 11 of 77