0x45 read one of the step's power and time values, 0x48 set the list file name, 0x49 read the list file name – B&K Precision 8500-8526 - Manual User Manual

Page 64: 0x4a set the memory partitioning for storing lists

Advertising
background image

0x45 Read one of the step's power and time values

Byte

offset

Meaning

3 to 4

2 byte little-endian integer specifying which step number in the list

5 to 8

4 byte little-endian integer specifying the power in units of 1 mW

9 to 10

2 byte little-endian integer specifying the step timing in units of 0.1 ms

11-24

Reserved

0x46 Set one of the step's resistance and time values

Byte

offset

Meaning

3 to 4

2 byte little-endian integer specifying which step number in the list

5 to 8

4 byte little-endian integer specifying the resistance in units of 1 m

9 to 10

2 byte little-endian integer specifying the step timing in units of 0.1 ms

11-24

Reserved

0x47 Read one of the step's resistance and time values

Byte

offset

Meaning

3 to 4

2 byte little-endian integer specifying which step number in the list

5 to 8

4 byte little-endian integer specifying the resistance in units of 1 m

9 to 10

2 byte little-endian integer specifying the step timing in units of 0.1 ms

11-24

Reserved

0x48 Set the list file name

Byte

offset

Meaning

3 to 12

List file name (ASCII characters)

13-24

Reserved

0x49 Read the list file name

Byte

offset

Meaning

3 to 12

List file name (ASCII characters)

13-24

Reserved

0x4A Set the memory partitioning for storing lists

8500 DC Load Series

Version: 030614

Page 64 of 77