Toggled transient operation – B&K Precision 8500-8526 - Manual User Manual

Page 37

Advertising
background image

Keys

Display

this width is ignored, so enter any convenient value.

Enter

Displays

LEVEL B= X.XXXA

. You are being prompted for the second

current value. Press the 1 and 0 keys to enter 10 A.

Enter

Enters the 10 A value, then displays

WIDTH B = X.XMS

. You're being

prompted for the duration of the 10 A load -- press the 1 and 0 key for 10
ms.

Enter

Enters the 10 ms duration, then displays

:CONTINUOUS

,

:PULSE

, or

:TOGGLED

. Use the

△ or ▽ arrow keys to display

:PULSE

.

Enter

Returns to the standard display of voltage and current.

To activate this pulse transient condition, press Shift + Tran, then press On/Off. The load will start
at the A level of current you entered (5 A). Press Shift + Trigger to cause the load to switch to the
B level of current (10 A). The 10 A current will last for the designated 10 ms, then switch back to the
A level (5 A).

NOTE: This example requires the :CONFIG:TRIGGER menu item be set to

:IMMEDIATE. If set

to :EXTERNAL, you'd trigger the instrument with a high TTL level signal on the trigger terminals on
the rear panel. If set to :BUS, you'd trigger the instrument with a software trigger.

Toggled transient operation

In toggled transient operation, the load starts at the stored parameter for the mode. When a trigger
is received, the load switches to the B value. When another trigger is received, the load switches to
the A value. It stays at the A value until another trigger is received, at which point it switches to the
B value. Further triggers cause the load to switch between the A and B values. Here's an example:

10A

5A

TRG

TRG

Toggled Transient Operation

Here are the keystrokes needed to setup this example:

Keys

Display

I-set

Standard display of voltage and current. Enter a value of 5 A (you may
need to press the I-set key again). Press Enter.

Enter

Enters the CC mode's value. The standard display is shown.

Shift + S-Tran

LEVEL A= X.XXXA

You are being prompted for the first current value.

Press the 5 key for 5 A.

Enter

Enters the 5 A value, then displays

WIDTH A = X.XMS

. You're being

8500 DC Load Series

Version: 030614

Page 37 of 77