Instrument won't turn on

If the instrument won't turn on when the POWER switch is pressed in, please ensure the power cord
is plugged into the back of the instrument and the other end of the cord is plugged into a live AC
power outlet.
If the instrument still won't turn on, remove the power cord from the instrument. Open the fuse
container on the rear panel and check the fuse for continuity.

If the fuse is an open circuit, replace it with the proper fuse indicated in the following table:

Model

Fuse specification for 110 VAC

operation

Fuse specification for 220 VAC

operation

8500

T0.5A, 250 VAC

T0.3A, 250 VAC

8502

T0.5A, 250 VAC

T0.3A, 250 VAC

8510

T1.25A, 250 VAC

T650 mA, 250 VAC

8512

T2.5A, 250 VAC

T1.25A, 250 VAC

8514

T2.5A, 250 VAC

T1.25A, 250 VAC

8518

T2.5A, 250 VAC

T1.25A, 250 VAC

8520

T2.5A, 250 VAC

T1.25A, 250 VAC

8522

T2.5A, 250 VAC

T1.25A, 250 VAC

8524

T5A, 250 VAC

T2.5A, 250VAC (For 8524-220V)

8526

T5A, 250 VAC

T2.5A, 250VAC (For 8526-220V)

WARNING

Ensure that the fuse is rated for 250 VAC operation. Lower voltage
rated fuses are not suitable for this instrument.

Error messages during power-up

EEPROM ERROR means the calibration data have been lost or the EEPROM is not functional.
ERROR CAL.DATA means that the calibration data have been lost.
If you experience these error messages, the instrument should be returned to B&K Precision for
service.

Keypad unlock procedure (Keypad was accidentally locked)

If you accidentally locked the keyboard, follow this procedure to unlock: The procedure below shows
how to unlock a model 8500 using “8512” as the unlock key.

1) Press Shift + Menu. You will see PASSWORD: displayed on the screen. Enter 8512, then

press the Enter key. Now you will see Connect on the display

2) Press Enter, locate menu Key Lock set, press Enter. Enter 8512, confirm with Enter, press