0x38 set cr mode transient resistance and timing, 0x39 read cr mode transient parameters, 0x3a select the list operation (cc/cv/cw/cr) – B&K Precision 8500-8526 - Manual User Manual

Page 61

Advertising
background image

Byte

offset

Meaning

15

Transient operation:

0 is CONTINUOUS
1 is PULSE
2 is TOGGLED

16-24

Reserved

0x38 Set CR mode transient resistance and timing

Byte

offset

Meaning

3 to 6

Value A of resistance in units of 1 m

. Little-endian 4 byte number.

7 to 8

Time for A resistance in units of 0.1 ms. Little-endian 2 byte number.

9 to 12

Value B of resistance in units of 1 m

. Little-endian 4 byte number.

13 to 14 Time for B resistance in units of 0.1 ms. Little-endian 2 byte number.

15

Transient operation:

0 is CONTINUOUS
1 is PULSE
2 is TOGGLED

16-24

Reserved

0x39 Read CR mode transient parameters

Byte

offset

Meaning

3 to 6

Value A of resistance in units of 1 m

. Little-endian 4 byte number.

7 to 8

Time for A resistance in units of 0.1 ms. Little-endian 2 byte number.

9 to 12

Value B of resistance in units of 1 m

. Little-endian 4 byte number.

13 to 14 Time for B resistance in units of 0.1 ms. Little-endian 2 byte number.

15

Transient operation:

0 is CONTINUOUS
1 is PULSE
2 is TOGGLED

16-24

Reserved

0x3A Select the list operation (CC/CV/CW/CR)

Byte

offset

Meaning

3

List operation mode:

0 is constant current (CC)
1 is constant voltage (CV)
2 is constant power (CW)

8500 DC Load Series

Version: 030614

Page 61 of 77