Test files – B&K Precision 8500-8526 - Manual User Manual

Page 40

Advertising
background image

Keys

Display

Esc Esc

Standard display

First, press the On/Off key. The load will sink the current that is stored as the CC mode parameter
(set it to 0 A if you don't want an initial current). Then press Shift + Trigger to have the list start
executing.
Should you wish to have the list only execute once after the trigger, you can edit the list to use

:ONCE

instead of

:REPEAT

.

To exit List mode press the Shift and Trigger key

Test files

Test files are a generalization of lists -- they let you generate a sequence of tests using different
modes, mode parameters, and durations. They are useful for executing a set of tests on a device,
then displaying whether the tests passed or failed. We will illustrate how to use test files by a short
example.
Suppose we have a small AC to DC power supply (a "wall-wart") and we want to set up an
acceptance test for a number of these devices. Our test will consist of two steps:

1. Set the DC load to constant current mode to draw the rated current of 0.35 A from the

device. The output voltage of the device at the rated current must be between 4.4 volts and
4.6 volts.

2. When the device operates into a short, the supplied current must be larger than 2.0 A.

Keys

Display

Shift + Menu

:CONFIG

▽ ▽

:LIST SET

Enter

:MODE SET

▽ ▽ ▽ ▽

:EDIT TEST FILE

Enter

MAX CURR= 3.000A

Set the maximum current to 2.5 A

2.5 Enter

MAX VOLT= 18.00V

Set the maximum voltage to 5 V

5 Enter

MAX POWER= 150.00W

Set the maximum power to 15 W

15 Enter

TEST COUNT= 6

Our test will consist of two steps, so we enter 2. You

can enter up to 20 steps.

2 Enter

CONST CURRENT

Since our first test is in constant current mode, we'll

just press Enter to select this mode.

Enter

SET 1= 0.210A

This is the prompt for the constant current value. We set

it to 0.35 A.

.35 Enter

SHORT OFF

We can turn a short on or off at this step. Here, we want

short off, so we just press Enter.

Enter

READBACK V

We're being prompted for what parameter to read back

and check. We want voltage, so we press Enter.

8500 DC Load Series

Version: 030614

Page 40 of 77