14 - TROUBLESHOOTING

14.1 - Malfunctions

[L] = Linux models
[W] = Windows models
[R] = Radio models
[C] = Cable models

The computer does not turn on

No power supply

Check the mains socket and the connections

Press the computer power-on button

see par. 1.1

[L] Reset button not pressed

Press the Reset button

see paragraph 1.1

Display off

Turn on the display (button on display)

The measuring head does not turn on or turns off unexpectedly

[R] Batteries fl at

Charge the batteries

[C] No power supply

Check the mains socket and the connections

[R] Battery recharging does not start

No power supply

Check the mains socket and the connections

Press the battery recharge button

see paragraph. 1.3

[R] The battery life is too short

Low ambient temperature

Normal phenomenon

Recharging not complete

Recharge the batteries for at least 5 hours

The battery has come to the end
of its lifetime

Replace the battery