Advanced parameters and commands

Pearleye Technical Manual V2.3.0

46

Advanced parameters and commands

This chapter describes the advanced configuration of the Pearleye P-007 LWIR
/ Pearleye P-030 LWIR models.

Two-point correction (A, B, E, J, K, N)

To configure the two-point correction, use the following parameters:

The parameter

E controls the operation mode of the two-point correction. The

following value assignments are valid:

Note

In most cases the factory setup of the camera is sufficient.

Only in special cases you need the advanced parameters and
commands.

Parameter Description

A

File number of the first correction image recorded at cold
reference temperature.

B

File number of the first correction image recorded at warm
reference temperature.

E

Two-point correction operating mode

J

Set value of the correction image of the actually chosen
correction data set recorded at cold reference temperature.

K

Set value of the correction image of the actually chosen
correction data set recorded at warm reference temperature.

N

File number for the set values of all correction data sets.

Table 17: Advanced parameters: two-point correction

Value

Description

0

Deactivated

Data is passed through transparently.

1

Activated

Correction data

A and B with the set values J and K of the

actual correction memory page

S is used.

Table 18: Parameter

E: values