LTP-Director Manual

12

Setting up an USB-ME256-System with multiple instances of LTP-Director

The scheme below shows the cable connections for setting up an USB-ME256 data acquisition
device with four MEA1060-BC amplifiers and one STG4008. When using a USB-ME128 and only
two amplifiers, the connections are similar. With the shown hardware configuration, you can
then use multiple instances of LTP-Director on the same computer to control each amplifier as
a completely independent unit. Please make sure to have four LTP-Director licenses to operate
the USB-ME256 data acquisition device with four MEA1060-BC amplifiers. Please see also chapter
“Installing the LTP-Director Software”.

Important: The filter amplifier (FA), which must be located between the MEA1060-BC
headstage and the USB-ME256 Analog In, is not shown in the scheme or the tables!

Important: After setting up the complete system with multiple instances of LTP-Director,
please check the gain settings of each single instance! May be the gain setting (1100)
is correct for the first instance, but incorrect for the following instances, when starting
the system for the first time.

The following tables show the same information as the scheme above.