LTP-Director

55

5.2.8

Recording Check List

1.

Close MC_Rack, MC_Stimulus and MEA_Select, if necessary.

2.

Open LTP-Director.

3.

Test Mode:

 Select Hardware (Settings / Hardware).

 Select electrode layout, MEA BC-Port and STG, if necessary.

 Select Data Path and Filename (Settings / Options).

 Load slice image, fit the image to the electrode grid.

 Adjust stimulation parameters.

 Select stimulation electrodes for pathways S1 and S2.

 Select up to four dendritic and / or somatic electrodes for online analysis.

 Switch to Measurement Mode.

4.

Measurement Mode:

 Optional: Open template.

 Fill in Lab Book.

 Run IO Curve.

 Optional: Save IO curves as pictures or ASCII files.

 Adjust stimulation parameters.

 Adjust detection levels.

 Select / Deselect experimental phases.

 Adjust phase parameters.

 Recheck stimulation pulse interval. (For testing, usually intervals of

5 or 10 s are used. For the real experiment, this might be too fast.)

 Recheck file name and file path.

 Optional: Open Experiment Comment window to make notes during

the experiment.

 Press Start.