Multichannel Systems STG4002 Manual User Manual

Page 34

Advertising
background image

Stimulus Generator 4002 Manual

30

4.

On the Settings menu, click Auto Sync to enable the Auto Sync feature. The Sync Out protocols

will be autocreated in synchrony with the pulse protocols on the assigned channels, will show up
in the WYSIWYG window immediately, and will be downloaded onto the STG with your next
download. (Please make sure that the display and download options are appropriate, that is,
all necessary Sync Out channels are selected.) The worksheets of the Sync Out channels assigned
to the Auto Sync feature will be unavailable, that is, you cannot edit them manually. The
following screen shot shows an example of a monophasic 100 μs voltage pulse on channel 1
and the corresponding Auto Sync pulse (Pre Time 0 μs, Post Time 100 μs) on Sync Out 1.

5.

If you later choose to edit the Sync Out worksheets of the active file manually, you can deselect

the Auto Sync feature on the Settings menu. Any information in the Sync Out worksheets that
was there before the Auto Sync option was enabled will be restored; and all worksheets will
be available for manual editing.

Advertising