Analog Output Signals

83

Example 3b: Current Mode: The green trace shows the current output (+I) with a load resistance
of 1 M

; and the pink trace shows the TTL output from the Sync Out that was used for

triggering the oscilloscope

Rise time with programmed 5 V and 500 μA pulses, 1 M

 load resistance.

Rise time (10 % / 90 %) = 120 μs

Result of Example 3: With a higher load resistance, the current pulse shows a different
kinetic behavior: The rise time increases significantly (about 100 μs). Please keep in mind
that the resistance of the oscilloscope probe with 10 M

 is parallel to the 1 M load resistance.

That means, that the maximum of the load resistance is reduced to 900 k

 only.