Downloading Stimuli

8.1

Downloading Stimulus Files

After having set up the stimulus file, you can download the file to the connected STG.
You can select the channels that you want to download or download the complete file.

You can start the stimulation directly after the download, or start it manually with the software
or hardware controls, or start it on a trigger. See also chapter "Operating the STG" for more
information.

Warning: Check the file thoroughly before downloading it on the STG. The stimulus display
is independent from the worksheet and from the download. Verify that you download

only

the channels that you want to use. Do

not

start the STG if you are not sure about the channel

configuration or the nature of the downloaded file.

1.

Select the channels for downloading under Download Channel and Download Sync.


2.

On the STG menu, click Download or click

on the toolbar to download only the selected

channels onto the STG or Download and Start

to download only the selected channels

and start the stimulation directly after the complete download.

Disable / enable warning messages

Cancel this option to disable the message boxes displaying compiler warnings during stimulus
download.

The two error messages that can be disabled are: "Warning: signal length shorter than the
maximum" and "Warning: non zero last amplitude value". They may present a nuisance
If you download multiple stimulus files in batch mode.

 On the Settings menu, select or deselect Enable Compiler Warnings.