Multichannel Systems STG4002 Manual User Manual
Page 91
Advertising
Analog Output Signals
87
Example 2b: Current pulse with programmed 16 μA for 20 μs and 20 μs interval and 10 k
load resistance.
The pulse is biphasic,16 μA. Green trace = current output (+I), pink trace = TTL trigger output
from the Sync Out.
Result of Example 2: The current output signal is delivered unmodified without any influence
of the length of the pulse, and an amplitude of 16 μA which is a hundredth of the limit of the
maximum capacity of the STG 4002