46-648 Jupiter Magnetostrictive Transmitters

17

1.6.4.4 Measurement Type: Level & Interface

Display

Action

Comment

1

Status
Level
% Output
Loop

Transmitter Display

Transmitter default display. Status, Level, % Output, and Loop
values cycle every 8 seconds.

2

Level

xxx.xx

Transmitter Display

Level measurement in centimeters or inches

3

% Output

xx.xx%

Transmitter Display

Level as a percentage of loop current span

4

Loop

xx.xx mA

Transmitter Display

Loop current output (mA)

5

IfcLevel

Transmitter Display

Displays interface level

6

MeasType

Select type of measurement

Choose Lvl&Ifc

7

Units

Select units of length

Choose cm or in

8

Probe Ln

xxx.x

Enter exact length of probe

6–420 inches (15.24–1066.8 cm)

9

Set 4mA

xxx.xxlu

Enter the PV value for 4 mA
point

Enter 4 mA point in level units

10

Set 20mA
xxx.xxlu

Enter the PV value for 20 mA
point

Enter 20 mA point in level units

11

Lvl Ofst
xxx.xxlu

Enter the level offset value

Changes zero level as referenced from probe tip

12

Damping

xx.x s

Enter damping filter time

0–1 second in 0.1 increments
1–25 seconds in 1.0 increments

13

Fault

22mA

Select loop current under
fault condition

3.6 mA, 22 mA or Hold

14

Poll Adr

xx

Enter HART polling address
number

0–15

15

Trim 4

xxxx

Adjust 4 mA point

Attach a meter to the output. If the output does not equal 4 mA,
adjust the value of the display until the meter reads 4 mA

16

Trim 20

xxxx

Adjust 20 mA point

Attach a meter to the output. If the output does not equal 20 mA,
adjust the value of the display until the meter reads 20 mA

17

Loop Tst

Enter a mA value

Set mA output to a value between 3.6 and 22.0 mA

18

DeadBand

xx.x

Factory Setting

Diagnostic, factory setting

19

Snsr Mnt

Enter mounting type

MLI Top (external probe; transmitter top mounted)
MLI Bot (external probe; transmitter bottom mounted)
Dir Near (NPT, BSP, and 600# or less flanged probe)
Dir Ext (flanged probes 900# class and over)

20

Trim Lvl

xx.xx

Enter value to adjust Level
reading

-20.00 inches

≤ Trim Lvl ≤ 20.00 inches

21

Trim Ifc

xx.xx

Enter value to adjust Interface
reading

-20.00 inches

≤ Trim Ifc ≤ 20.00 inches

22

F1 Cnts

Display only

For factory diagnostic use

23

F2 Cnts

Display only

For factory diagnostic use