7 specifications, 1 performance – Orion Instruments Jupiter 2xx User Manual

Page 34

Advertising
background image

34

46-648 Jupiter Magnetostrictive Transmitters

2.7

Specifications

2.7.1 Performance

Accuracy

±0.015"

Repeatability

±0.005% of full span or 0.005" (0.127 mm) (whichever is greater)

Linearity

0.020% of full span or 0.031" (0.794 mm) (whichever is greater)

Maximum level rate of change

6 inches per second (models with HART)

Response time

0.1 second

Warm-up

<5 second

Upper dead zone

None

Lower dead zone

<2" (5 cm), SIL 2: <5" (13 cm)

Ambient temperature range

Transmitter:

-40° to +175° F (-40° to +80° C)

LCD:

-10° to +160° F (-20° to +70° C)

Process temperature

External Mount:

-40° to +248° F (-40° to +120° C)

-320° to +850° F (-195° to +455° C) (with factory insulated MLI)

Direct Insertion:

-40° to +200° F (-40° to +95° C)

-40° to +500° F (-40° to +260° C) (high temperature probe)

Humidity

0 to 99% non-condensing

Electromagnetic compliance

EN 61326

Environmental protection compliance

EN 60654-1

Drop protection compliance

EN 50178

Surge Protection Compliance

EN 61326 (1000 V)

Maximum Pressure (Direct Insertion)

1700 psig @ +100° F (117 bar @ +38° C)

(limited to the pressure rating of the selected flange or float)

Advertising