Orion Instruments Jupiter 2xx User Manual

Page 28

Advertising
background image

28

46-648 Jupiter Magnetostrictive Transmitters

Check Loop

Current at Terminal

Board

Loop Current

Above 23 mA?

Remove Electronic

Module

Loop Current

Above 1 mA?

Replace Terminal

Board

Replace Electronic

Module

Yes

Yes

No

No

Failure to Operate

with Correct Loop

Inputs

Display

Working?

Yes

No

Retest Unit

Good

Display

Key Board

Responds?

Replace Electronic

Module

No

Check Loop

Current at Wiring

Board

Loop Current

22 - 4.0 mA?

Adjust 4 and 20 mA

Unit has

HART?

HART

Working?

Yes

Yes

No

Yes

No

Address

Correct?

Correct Address

Replace Electronic

Module

No

Yes

Unit with Display

Communicates

No

Yes

Reseat Boards

Display

Working?

Retest

No

Yes

1

2

4

Display

Working?

Yes

Retest Unit

continued on next page

¯