Fairbanks 2800 Series Intrinsically Safe Instrument User Manual
Page 40
Advertising
Section 3: Programming
01/12
40
50647 Rev. 9
NOTE:
The 4-20mA Output is enabled by selecting
AnALog
in
Port A.
PROM Version 6.2+ Notes
The 4-20mA output remains locked on the last reading at the time the
PGR
was pressed.
An Overload (
oL
) condition produces a reading of 24mA.
An Underload (
UL
) condition produces a reading of 3.5mA.
While in the Sleep mode, the 4-20mA produces a reading of 3.5mA, and it is
not tracked until the instrument is re-activated.
MENU ITEM
DESCRIPTION
4-20
─
─
grLooP
: Tracks the GROSS Weight.
─
ntLooP
: Tracks the NET Weight.
C A U T I O N
The 4-20ma accessory is a passive device!
The power is to be supplied by others, and MUST
BE
a
DEDICATED
and an
ISOLATED
power supply.