Appendix i: troubleshooting/error codes – Fairbanks 2800 Series Intrinsically Safe Instrument User Manual

Page 49

Advertising
background image

01/12

49

50647 Rev. 9

Appendix I: Troubleshooting/Error Codes




FAULT

CAUSE

SOLUTION

-UL--

LOADCELL FAILURE

Scale restricted.
Debris under scale.

Scale behind 0.

Load Cell failure.

Correct the bind or restriction.
Remove debris under the
scale/cells.
Replace items removed from
the scale.
Call Fairbanks service.

--oL--

LOADCELL

FAILURE

Overloaded Scale.
Load Cell failure.

Remove weight from the scale.
Call Fairbanks service.

RECHARGE

Battery voltage low
between 5.4 to 5.32VDC.

Recharge the battery.

LoBAtt

Battery voltage low, between
5.3

to 5.12VDC

Recharge the battery
IMMEDIATELY.

err 069

Programming input error.

Recycle the power to
instrument.

likely lc error

0shift

out of range


Load Cell Failure.

Remove weight from the scale.
Re-zero the scale.
Call Fairbanks service.