Levenhuk C800 NG Digital Camera, USB 2.0 User Manual

Page 136

Advertising
background image

Help Manual

128

middle of the microscope field and try to find the ruler clearly. Set the

Unit

to

Pixel

and the video

Resolution

to the maximum one (640X480 0.35M camera) and

Zoom

ratio to 100% (see

);

5. Choosing

Options>Calibrate•••

or clicking “

” on the toolbar, a red line with

pixels number and 0.000um length is overlaid on the video. At the same time, a

dialog called

Calibrate

will be displayed over the

Video

window;

6. Try to align the two ends of the red line with the video ruler scale (Try to pull the

red line as long as possible to keep the calibration accuracy);

7. Enter or select the current microscope objective lens

Magnification

in the

Magnification

field. The current

Magnification

is 10X;

8. Read the overlaid (by the red line) micro-ruler actual length and fill it in the

Actual

Length

field, the current

Actual Length

is 450um. This

Actual Length

will also be

displayed in the middle of the red line just behind the pixel number. At the same time,