9.7.2 Rotate to zero

If the

Video Watermark

was rotated, the

Rotate to zero

menu will be enabled.

Choosing this menu will rotate the

Video Watermark

to zero degree.

9.8 Gray Calibration•••

This function can make specified area image brightness a desired value among
various scenarios, achieving the continuity requirement of the observation. The

Gray

Calibration

steps are summarized as follows:

1. Click

Exposure & Gain

on the

Camera

sidebar to extend the

Exposure & Gain

group,

uncheck

Auto Exposure

box.

2. Choose

Edit>Image Select

command or click

on the toolbar to select a

reference region, and choose “

Setup>Gray Calibration

” command. A dialog called

Gray Calibration

will be brought up to display the current ROI

Average Gray

. Now the

brightness of the microscope light source can be adjusted until the

Average Gray

closing to the desired value. Click “

OK

” to finish the calibration and return to the

application area. The desired value here is around is 200.