Specifications affected by electronic attenuator – Agilent Technologies N9010A User Manual

Page 116

Advertising
background image

116

Chapter 8

Option EA3 - Electronic Attenuator, 3.6 GHz

Specifications Affected by Electronic Attenuator

Specifications Affected by Electronic Attenuator

Specification Name

Information

Frequency Range

See

“Range (Frequency and Attenuation)” on page 117

.

1 dB Gain Compression Point

See

“Distortions and Noise” on page 118

.

Displayed Average Noise Level

See

“Distortions and Noise” on page 118

.

Frequency Response

See

“Frequency Response” on page 119

.

Attenuator Switching Uncertainty

The recommended operation of the electronic attenuator is with
the reference setting (10 dB) of the mechanical attenuator. In
this operating condition, the Attenuator Switching Uncertainty
specification of the mechanical attenuator in the core
specifications does not apply, and any switching uncertainty of
the electronic attenuator is included within the

“Electronic

Attenuator Switching Uncertainty” on page 120

.

Absolute Amplitude Accuracy,

See .

“Absolute Amplitude Accuracy” on page 119

.

Second Harmonic Distortion

See

“Distortions and Noise” on page 118

.

Third Order Intermodulation
Distortion

See

“Distortions and Noise” on page 118

.