Other microwave preselector bypass specifications – Agilent Technologies N9010A User Manual

Page 137

Advertising
background image

Chapter 12

137

Option MPB - Microwave Preselector Bypass

Other Microwave Preselector Bypass Specifications

Other Microwave Preselector Bypass Specifications

Description

Specifications

Supplemental Information

Frequency Response

(Maximum error relative to
reference condition (50 MHz)

Swept operation

a

, Attenuation

10 dB)

a. For Sweep Type = FFT, add the RF flatness errors of this table to the IF Frequency Response errors. An

additional error source, the error in switching between swept and FFT sweep types, is nominally
±0.01 dB and is included within the “Absolute Amplitude Error” specifications.

Refer to the footnote for

Band Overlaps on page 21

.

Modes above 18 GHz

b

b. Signal frequencies between 18 and 26.5 GHz are prone to additional response errors due to modes in

the Type-N connector used with frequency Option 526. With the use Type-N to APC 3.5 mm adapter
part number 1250-1744, there are nominally six such modes. The effect of these modes with this con-
nector are included within these specifications.

20 to 30

°C

Full Range

95

th

Percentile (

2σ)

3.5 to 8.4 GHz

±0.9 dB

±1.5 dB

±0.42 dB

8.3 to 13.6 GHz

±1.0 dB

±2.0 dB

±0.50 dB

13.5 to 17.1 GHz

±1.3 dB

±2.0 dB

±0.50 dB

17.0 to 22.0 GHz

±1.3 dB

±2.0 dB

±0.53 dB

22.0 to 26.5 GHz

±2.0 dB

±2.8 dB

±0.66 dB

26.4 to 34.5 GHz

±2.0 dB

±3.0 dB

±0.80 dB

34.4 to 44 GHz

±3.1 dB

±4.8 dB

±1.21 dB

Advertising