Lo output – Agilent Technologies N9010A User Manual

Page 134

Advertising
background image

134

Chapter 11

Option EXM - External Mixing

Other External Mixing Specifications

c. The amplitude accuracy of a measurement includes this term and the accuracy with which the settings of

corrections model the loss of the external mixer.

Description

Specifications

Supplemental
Information

LO Output

Frequency Range

3.75 to 14.0 GHz

Output Power

a

20 to 30

°C

Full Range

3.75 to 7.0 GHz

b

+15.0 to 18.0 dBm

+14.5 to 18.5 dBm

7.0 to 8.72 GHz

b

+15.0 to 18.0 dBm

+13.5 to 18.8 dBm

7.8 to 14.0 GHz

c

+14.0 to 18.5 dBm

Not specified

Second Harmonic

–20 dB (nominal)

Fundamental Feedthrough
and Undesired Harmonics

c

–15 dB (nominal)

VSWR

<2.2:1 (nominal)

a. The LO output port power is compatible with Agilent M1970 and 11970 Series mixers except for the

11970K. The power is specified at the connector. Cable loss will affect the power available at the
mixer. With non-Agilent mixer units, supplied loss calibration data may be valid only at a specified LO
power that may differ from the power available at the mixer. In such cases, additional uncertainties
apply.

b. LO Doubler = Off settings.
c. LO Doubler = On setting. Fundamental frequency = 3.9 to 7.0 GHz.

Advertising