Power control – Agilent Technologies N9010A User Manual

Page 296

Advertising
background image

296

Chapter 33

W-CDMA Measurement Application

Measurements

Description

Specifications

Supplemental Information

Power Control

Absolute power measurement

Using 5 MHz resolution bandwidth

Accuracy

0 to

−20 dBm

±0.7 dB (nominal)

−20 to −60 dBm

±1.0 dB (nominal)

Relative power measurement

Accuracy

Step range

±1.5 dB

±0.1 dB (nominal)

Step range

±3.0 dB

±0.15 dB (nominal)

Step range

±4.5 dB

±0.2 dB (nominal)

Step range

±26.0 dB

±0.3 dB (nominal)

Advertising