Electronic attenuator switching uncertainty, Electronic – Agilent Technologies N9010A User Manual

Page 120

Advertising
background image

120

Chapter 8

Option EA3 - Electronic Attenuator, 3.6 GHz

Other Electronic Attenuator Specifications

Description

Specifications

Supplemental Information

Electronic Attenuator Switching
Uncertainty

(Error relative to reference
condition: 50 MHz, 10 dB
mechanical attenuation, 10 dB
electronic attenuation)

Attenuation = 0 to 24 dB

9 kHz to 3.6 GHz

See note

a

a. The specification is

±0.14 dB. Note that this small relative uncertainty does not apply in estimating

absolute amplitude accuracy. It is included within the absolute amplitude accuracy for measurements
done with the electronic attenuator. (Measurements made without the electronic attenuator are treated
differently; the absolute amplitude accuracy specification for these measurements does not include
attenuator switching uncertainty.)

Advertising