Agilent EXA Signal Analyzer

Dynamic Range

13.6 to 26.5 GHz

x

+7 dBm

13.5 to 17.1 GHz

x

+9 dBm

17.0 to 26.5 GHz

x

+8 dBm

a. See the IF Prefilter Bandwidth table in the Gain Compression specifications on

page 43

. When the tone

separation condition is met, the effect on TOI of the setting of IF Gain is negligible. TOI is verified
with IF Gain set to its best case condition, which is IF Gain = Low.

b. TOI is verified with two tones, each at

−16 dBm at the mixer, spaced by 100 kHz.

c. Intercept = TOI = third order intercept. The TOI is given by the mixer tone level (in dBm) minus (dis-

tortion/2) where distortion is the relative level of the distortion tones in dBc.

Nominal Dynamic Range vs. Offset Frequency vs. RBW for Freq Option

526 [Plot]