8.7 HOLD

The FLXA21 has a function to hold the mA output at a preset value (default: “Last”).
Use this menu to hold the output.

For the settings, see “n Configure Hold” on page 7-8.
During commissioning or quick setup, the output is automatically held.
Setting “Hold during Calibration” to “Disabled” deactivates the hold function during calibration.

Press

to select Execute: HOLD and then choose Manual Hold ON or Manual Hold OFF.

This allows you to set up manual hold.

10.00

Tag:FLXA21–SC

25.0

4mA

20mA

Conduct1-TC1

°C

mS/cm

HOLD

Figure 8.8

Example of the display with the manual hold enabled

To cancel manual hold, press the lit

HOLD

section on the Main display.

8.8

Temporary output

If measurement cannot be made due to replacement of one of the sensors, assignment of mA

output can be temporarily changed from the currently setup sensor to the other sensor. This

feature is enabled only when two sensors are connected.

Press

to select Execute: Temporary output and then choose the process parameter to be

assigned on the Temporary output screen.

When process parameter assignment has been modified, the indication at the upper left of the

screen changes to .
The feature is disabled the moment the screen returns to the Main display, and the setting

(process parameter in mA output setting) returns to the original condition.