<10. COMMISSIONING OF ISC (Inductive Conductivity)>

10-4

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

l

Process temperature compensation

Select a temperature compensation method from among None, TC, NaCl, and Matrix. “None”

does not perform the temperature compensation.

l

TC

This method uses the linear compensation function.
For how to calculate a temperature coefficient of the compensation function, see “l Configure

calculated temperature coefficient (TC).” on page App.3-1.

l

NaCl

This method uses the standard temperature compensation function with NaCl solution.
For details, see “l NaCl or standard temperature compensation” on page App.3-1.

l

Matrix

This method uses the temperature compensation matrix, which is a table of the conductivity

values at different temperatures and concentrations for a specific solution. The measured

conductivity is compensated based on this matrix to the conductivity at the reference

temperature.
Prepared matrixes and user-defined matrixes can be used. Two different user-defined

matrixes are available.
For details, see “l Matrix compensation” on page App.3-3.

When the measured temperature or the precompensated conductivity is out of the range, the

temperature compensation error (warning) will be issued. This is not a device error.
For more details, see “■ Temperature compensation error” on App. 3-8.

Any two of these three methods can be set for conductivity measurement as “Temp.comp.1” and

“Temp.comp.2”. The result of either method can be output or displayed (see “10.2 Output setup”

or “10.6 Display setup”).
When “Measurement” is set to “Conduct. + Concentr.” for the Measurement, only one method

can be set.

NOTE

The temperature compensation is not performed around zero. In this case, a warning may be

issued.

10.1.5

Calibration settings

The screen flow differs depending on the combination of objects to be measured.

l

Air adjust limit

Generally, air calibration is not required. To avoid the effects of the cable on the measurement of

lower conductivities such as pure water, a “zero” calibration with a dry sensor may be done.

Perform a zero-calibration using a dry sensor to prevent the cable from affecting the

measurement. If a terminal box (BA20) and extension cable (WF10) are used, perform the zero-

calibration while they are connected.
As the calibration is performed in air, the resistivity is infinite (open connection). Higher

conductivity values than the air adjust limit indicate that the cell is not in air or is still wet. To

prevent wrong air calibrations, a limit must be given here.