<10. COMMISSIONING OF ISC (Inductive Conductivity)>

10-3

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

10.1.1 Measurement

Select a measurement parameter from Conductivity, Concentration, and Conduct. + Concentr.

The setting of the measurement parameter changes the menu structure in Error configuration,

Display setup, etc. accordingly.

10.1.2

Configure sensor

This section describes how to configure the sensor to be connected.

l

Measuring unit

Either /cm or /m can be chosen here.
The process values will be expressed in S/cm or S/m.

l

Cell constant (factory)

The cell constant given at factory shipment is set.
When a new sensor is connected, set the new cell constant which is indicated on the sensor.
When a cell constant is changed, a calibrated cell constant (c.c. (adjusted)) will be the same

value of the new cell constant (factory) after resetting an old one.

10.1.3

Temperature settings

Select the temperature element used for compensation from among Pt1000 or NTC30k. Select

the same type as the temperature element that is actually connected.

Celsius (°C) or Fahrenheit (°F) temperature units are available. If the unit is changed, the

following values are also recalculated automatically to the new unit:

• Manual temp.
• Reference temp.
• Temp. coefficient
• Temp. ranges in the matrix

10.1.4

Temperature compensation

l

Temperature compensation

Two methods can be used: Automatic and Manual. Select Automatic when a temperature

element is used or Manual when a manually set temperature is used.

NOTE

When Manual is selected on the Temperature compensation, a process temperature should

be set in the “Manual temp.” A temperature shown on the Main display is this manually set

temperature.

l

Reference temperature

The temperature compensation function converts the measured conductivity value to that at the

reference temperature. The reference temperature can be changed. The default value is 25ºC.