<2. WIRING AND INSTALLATION>

2-10

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

2.5.1

Wiring the pH/ORP sensor

n

pH Measurement

Conventional pH sensors are connected to the module as follows:

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Glass (measure)
16 Shield
17 Shield

FLXA21

REF

TC

PH

Liquid

Earth

In addition to the wiring of the sensor, insure that a jumper for low-impedance sensor inputs is

installed. The jumpers can be found on the plastic sensor module cover and can be stored in the

lower level module wiring cover.

• pH Glass Electrode is a high impedance sensor input
• Standard reference electrodes and an ORP/REDOX electrode are low impedance sensor

inputs

• Special electrodes using 2 glass sensor (example: Pfaudler, SC24V) do not need jumpers.

Terminals 15-16 are identified as input 1 (High Impedance) and terminals 13-17 are defined as

input 2 (Low Impedance). For conventional pH sensors, the jumper is placed as illustrated:

16

15

19

17

13

18

14

12

11

PH

Input 1

Input 2

Glass sensor on Input 1.

Reference sensor on Input 2.