Calibration of ph/orp, Calibration of ph/orp -1, Ph 5 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 75

Advertising
background image

<5. CALIBRATION OF pH/ORP>

5-1

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

5.

CALIBRATION OF pH/ORP

Before pH measurement, calibrate the pH sensor with the standard solution.
Before ORP measurement, check the electrode as a part of regular maintenance.

NOTE

A default is “10 min” for “Auto Return”. When maintenance like a calibration that may take much

time is performed, “60 min” or “Disable” is recommended to be selected. (Refer to the section

4.7.3.)

Calibration

Execute:
Calibration
HOLD
Temporary output

Setup:
Commissioning
Change language
Start Quick Setup

Execute & Setup

Calibration pH

pH
ORP
Temperature

Manual
Automatic
Sample

Temperature calibration

or

Only when “pH + ORP” is selected
on the Sensor setup screen

Figure 5.1

Calibration

The calibration items include pH, ORP, rH, and Temperature.
The item to be calibrated is determined according to the settings made in the Sensor setup and

Measurement setup in Commissioning.

Press

to select Execute: Calibration, choose the item to be calibrated, configure the

calibration settings, then perform calibration.

NOTE

Note the following when performing calibration with buffer solutions.
1. Before starting a calibration, make sure the electrode system is properly cleaned and

the electrodes are fully functional. They must then be rinsed with clean water to avoid

contamination of the calibration solution(s).

2. Always use fresh buffer solution to avoid the risk of introducing errors from contaminated or

old solutions. Buffers supplied as liquids have a limited shelf life, especially alkaline buffers,

which absorb CO

2

from the air.

3. Yokogawa strongly recommends NIST/DIN 19266 buffer standards for the best accuracy

and buffer capacity. Commercially adjusted buffers (e.g., pH 7.00, 9.00 or 10.00) are a

compromise as a standard, and are often supplied without a temperature dependency

curve. Their stability will never be as good as NIST (JIS equivalent) solutions.

Always ensure that the sensors are properly conditioned, clean and filled with the correct

electrolyte solution (if appropriate) before starting a calibration. Refer to the sensor instructions

for details.

PH

5

Advertising
This manual is related to the following products: