3 installing the cable glands, Installing the cable glands -3, Caution – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 30: Ph sc isc do, Sencom

Advertising
background image

<2. WIRING AND INSTALLATION>

2-3

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

2.3

Installing the cable glands

The supplied cable glands are for cables with an outside diameter of 6 to 12 mm (0.24 to 0.47

inches). Unused cable entry holes must be sealed with cable glands including the supplied close

up plugs.

(A hole is drilled, if specified.)

Plastic Housing

Stainless Steel Housing

(For sensor 2 cable)

For sensor 1 cable

(For sensor 2 cable)

For sensor 1 cable

For power supply

For power supply

For grounding cable

F0202.ai

Figure 2.3

Cable gland diagram

CAUTION

Be careful not to be injured by the sharp hole edges on the housing.

Install the supplied cable gland as shown in Figure 2.4. When using an adapter for conduit work,

see Figure 2.5.

F0203.ai

Cable gland nut

O-ring

Gable gland

Sleeve

(for grounding cable line

of the plastic housing)

Plastic Housing

Cable gland cap

Cable gland nut

Gaskets

Washer

Stainless Steel Housing

Gable gland

Cable gland cap

Rubber plug

(without wiring)

Figure 2.4

Cable glands

The unused cable glands should be sealed with the supplied close up plug.

PH

SC

ISC

DO

SENCOM

2

Advertising