n

Sensor wellness

At the Sensor wellness window, the soundness of a module is displayed. A larger number of n

in each gauge indicates that the parameter concerned is sound. A gauge is indicated for only

those parameters whose sensor wellness setting is “enabled,” while a bar (----) is displayed if the

sensor wellness setting is “disabled.”
Sensor wellness setup can be made in Commissioning → Measurement setup → Sensor diag.

settings. For details, see section 7.1.7.
The “Reset wellness data” button can reset wellness data.
When a sensor is exchanged or replaced, sensor wellness data should be reset.

NOTE

When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to

the figure 6.8.)

n

Last calibrated

= date on which the last sensor calibration was performed. The displayed value of the Zero is the

result of this calibration. The displayed value of Slope was calibrated on this date only if the last

calibration was a 2-point calibration.

n

Calibration due

= the date when the calibration must be done next according to the settings of the calibration

interval. The calibration intervals are set in Commissioning → Measurement setup → Calibration

settings → Limits and timing.

n

Projected replacement

The projected maintenance function predicts the date when the sensor unit will need cleaning or

recalibrating for maintaining measurement accuracy. This function checks the cleanliness of the

sensor by monitoring the polarization of the electrodes every 24 hours and the cell constant after

the calibration. Clean the sensor before the predicted date.
The function predicts the following two dates:

When the cell constant crosses the upper or lower limits
When the value of the polarization of the electrodes reaches the alarm point (15% or more).

The function displays the earlier date with the status (the status is displayed in parentheses).
As shown in Figure 6.7, the date is predicted based on the intersection point of the upper or lower

limits and the extrapolated line of the values obtained by the least squares method.

Day

Value

Lower limit

Upper limit

Current

date

Projected

maintenance date

F0310.ai

Figure 6.7