<4. COMMISSIONING OF pH/ORP>

4-17

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

However, for applications where the pH during wash shows almost the same value as the normal

pH, the difference will be approximately zero. In such a case, disable the wash recovery check.

An example of such applications is the monitoring of waste water pH. If you select water jet

cleaning in such a case, the normal pH as well as the washing time pH will be around pH 7; the

difference will be approximately zero, so the recovery time check will not work normally.
This function is available either for ORP measurement only.

pH

Time

Good Electrode

Bad Electrode

tR : Recovery Time

tW : Wash Time

tI : Wash Period

1/2∆pH

∆pH

1/3tR

tI : Wash Period

tW : Wash Time

tR : Recovery Time

F050605_3.ai

Figure 4.6

4.6.6

Factory setup

For “Factory setup,” there is no item to be set by the user.

NOTE

This menu is for service engineers only. This section is protected by a password. Attempting to

change data in the factory adjustment menu without the proper instructions and equipment could

result in corruption of the instrument setup and damage the performance of the unit.

4.7

Display setup

This screen is used to make various settings relating to screen display.

NOTE

Settable items differ depending on settings in “Sensor setup” and “Measurement setup.”