App index – Yokogawa µR20000 User Manual

Page 25

Advertising
background image

2-9

IM 04P01B01-17E

Using

the
Ethernet

Interface

(/C7
Option)

1

2

3

4

5

6

App

Index

8.

Set the user name and then press the key.

Key operations
• Use the

key to select the digit for entering a character.

• Use the

CHARACTER key to select the character type.

• Use the

key to select the character you wish to enter.

User=

9.

Set the password and then press the key in the same fashion as in step 8.

The setting complete screen appears.

Password=

Ethernet login

Setting complete

To register other users, press the key to return to step 6 and repeat steps 8, 8,
and 9.

Saving the Settings

1.

Press the

ESC key to return to the Basic= screen.

2.

Press the

key to select

End and then press the key.

3.

Press the

key to select

Store and then press the key.

The settings are activated, and the Operation mode screen appears.

Explanation

You can limit the users that can access the Setting/Measurement and Maintenance/Test
servers on the recorder via the Ethernet interface.

• Enabling/Disabling the Login Function

Set whether to use the login function.

• Registering Users

• User level

Select either of the user levels, administrator or user.

• Administrator

One administrator can be registered. An administrator has the authority to use all

Setting/Measurement server and Maintenance/Test server commands.

• User

Six users can be registered. A user has limited authority to use the commands. See

section 4.2.

• Selecting Whether to Register (On/Off) the User

If On is selected, set the user name and password.

• Setting the User Name

• Set the user name using up to 16 alphanumeric characters.
• The same user name can not be registered.
• Since the word “quit” is reserved as a command on the recorder, the user name

“quit” is not allowed.

• Setting the Password

Set the password using up to 4 alphanumeric characters and spaces.

2.5 Registering Users

* When the key, key, or

CHARACTER key is pressed while holding down the SHIFT key, the operation

is reversed as when the respective key is pressed by itself.

Advertising
This manual is related to the following products: