6 control commands, 6 control commands -28, Ds switches the execution mode – Yokogawa µR20000 User Manual

Page 71: Ps starts/stops recording, Ud switches the screen/switches the channel, Ak executes alarm acknowledge (alarm ack), Tl starts/stops/resets computation (/m1 option), Mp starts/stops manual print, Ms prints the message, Ac clears the alarm printout buffer

Advertising
background image

4-28

IM 04P01B01-17E

4.6 Control Commands

DS

Switches the execution mode.

Syntax

DS p1<terminator>
p1 Mode

0

Run mode

1

Basic Setting mode

Example Switch to Basic Setting mode.

DS1

Description The execution mode cannot be changed to Basic

Setting Mode while recording or computation is
in progress.

PS

Starts/Stops recording.

Syntax

PS p1<terminator>
p1 Start/Stop recording.

0

Start

1

Stop

Example Start recording.

PS0

UD

Switches the screen/switches

the channel.

Returning to the data display screen

Returns from the Setting mode screen, the screen that
appears when the FUNC key is pressed, or the screen that
appears when the DISP MENU key is pressed back to the
measurement data display screen.
Syntax

UD p1<terminator>
p1 Fixed to 0.

Example Return to the data display screen.

UD0

Displaying the specified data display screen

Syntax

UD p1,p2<terminator>
p1 Fixed to 1.
p2 Screen number (1 to 15)

Example Display data display screen 2.

UD1,2

Switching the display channel (manual switching)

Syntax

UD p1<terminator>
p1 Fixed to 2.

Example Switch the display channel.

UD2

Description • An error results if the specified screen is set to

SKIP for UD1.

• UD2 is valid on screens whose display

channel is set to manual switching.

AK

Executes alarm acknowledge

(Alarm ACK).

Syntax

AK p1<terminator>
p1 Executes alarm acknowledge (0)

Example Execute alarm acknowledge.

AK0

TL

Starts/stops/resets computation

(/M1 option).

Syntax

TL p1<terminator>
p1 Operation type

0

Computation start

1

Computation stop

2

Computation reset

Example Start the computation.

TL0

MP

Starts/Stops manual print.

Syntax

MP p1<terminator>
p1 Operation type

0

Manual print start

1

Manual print stop

Example Start manual print.

MP0

LS

Starts/Stops the list (setting

information) printout.

Syntax

LS p1<terminator>
p1 Operation type

0

List print start

1

List print stop

Example Start list print.

LS0

Description List print refers to printing of settings of Setting

mode.

SU

Starts/Stops the setup list (basic

setting information) printout.

Syntax

SU p1<terminator>
p1 Operation type

0

Setup list print start

1

Setup list print stop

Example Start setup list print.

SU0

Description Setup list print refers to printing of settings of

Basic Setting mode.

MS

Prints the message.

Syntax

MS p1<terminator>
p1 Message number (1 to 5)

Example Print the message of message number 3.

MS3

Description The message string is set with the SG command.

AC

Clears the alarm printout buffer.

Syntax

AC p1<terminator>
p1 Clear the alarm printout buffer (0)

Example Clear the alarm printout buffer.

AC0

4.6 Control Commands