4-29

MC

Clears the message printout

buffer.

Syntax

MC p1<terminator>
p1 Clear the message printout buffer (0)

Example Clear the message printout buffer.

MC0

VG

Resets the report data of the

periodic printout.

Syntax

VG p1<terminator>
p1 Operation type

2

Reset the report data of the periodic
printout.

Example Reset the report data of the periodic printout.

VG2

Description This setting is valid when the recorder is set to

print the report data using periodic printout.

YC

Initializes settings.

Syntax

YC p1<terminator>
p1 Initialization type

0

Initialize the Setting mode and Basic
Setting mode settings.

1

Initialize the Setting mode settings.

Example Initialize the Setting mode and Basic Setting

mode settings.
YC0

UY

Stops the record position

adjustment.

Syntax

UY p1<terminator>
p1 Fixed to 0.

Query

UY?

Example Stop the record position adjustment.

UY0

Description Returns the execution status of the record

position adjustment in response to a query.

0: Stopped, 1: In progress

4.7 Output Commands

BO

Sets the byte output order.

Syntax

BO p1<terminator>
p1 Byte order

0

Outputs the data MSB first.

1

Outputs the data LSB first.

Query

BO?

Example Output the data MSB first.

BO0

Description This command applies to the byte order of

numerical data during BINARY output.

CS

Sets the checksum (/C3 option).

Syntax

CS p1<terminator>
p1 Use/Not use the checksum

0

Not use

1

Use

Query

CS?

Example Use the checksum.

CS1

Description This command can be used only on the RS-

422A/485 communication interface.

IF

Sets the status filter.

Syntax

IF p1<terminator>
p1 Status filter value

(0.0.0.0 to 255.255.255.255)

Query

IF?

Example Set the status filter value to 1.0.4.0.

IF1.0.4.0

Description For details, see chapter 6.

Note

Initialization of BO/CS/IF Command Settings

• For serial communications

Settings entered using the BO/CS/IF commands revert to their

initial values when the recorder is reset (when the recorder is

power cycled, or the user exits the basic setting mode with the

YE command).

• Byte output order, checksum: 0

• Status filter: 255.255.255.255

If the recorder is reset, you must restore these settings.

• For Ethernet communications

Settings entered using the BO/IF commands revert to their

initial values when the connection to the recorder is cut. After

reconnecting the recorder, you must reenter the settings.