Flag, Block member, Special data value – Yokogawa µR20000 User Manual

Page 90

Advertising
background image

5-14

IM 04P01B01-17E

• Flag

The meanings of the flags are given on the table below. The flags are valid during

FIFO data output. The flags are undefined for other cases.

Bit

Flag

Flag

Meaning of the Flag

0

1

6

5

4

3

2

No

Yes

Indicates that the decimal position or unit information was

changed during measurement.

1

No

Yes

Indicates that the FIFO acquiring interval was changed during

measurement.

0

No

Yes

Indicates that the internal process took too much time

(computation, for example) and that FIFO dropout occurred.

The bits that have “–” for the flag column are not used. The value is undefined.

• Block Member

Name

BINARY Value

Year

0 to 99

Month

1 to 12

Day

1 to 31

Hour

0 to 23

Minute

0 to 59

Second

0 to 59

Millisecond

0 to 999

DST

1 (=summer) or 0 (=winter)

Measurement, computation

00H: measurement, 80H: computation

Channel

µR10000: 01 to 06 and 31 to 42
µR20000: 01 to 24, 31 to 54

Alarm status*

0 to 8

1 byte

1 byte

A2 (Level 2) A1 (Level 1) A4 (Level 4) A3 (Level 3)

A1 (Bit 0 to 3)
A2 (Bit 4 to 7)
A3 (Bit 0 to 3)
A4 (Bit 4 to 7)

Measured data

0 to FFFFH

Computed data

0 to FFFFFFFFH

* BINARY value 0 to 8 is entered in the upper and lower 4 bits of a byte (8 bits) for the alarm

status. The binary values 0 to 8 correspond to H (high limit alarm), L (low limit alarm), h
(difference high-limit alarm), l (difference low-limit alarm), R (high limit on rate-of-change
alarm), r (low limit on rate-of-change alarm), T (delay high limit alarm), and t (delay low limit
alarm) as follows:
0: no alarm, 1: H, 2: L, 3: h, 4: l, 5: R, 6: r, 7: T, and 8: t.

• Special Data Value

The measured/computed data take on the following values under special conditions.

Special Data Value

Measured Data

Computed Data

+ Over

7FFFH

7FFF7FFFH

– Over

8001H

80018001H

Skip

8002H

80028002H

Burnout (when “up” is set)

7FFAH

7FFF7FFFH

Burnout (when “down” is set)

8006H

80018001H

Error

8004H

80048004H

Undefined

8005H

80058005H

Note

The number of blocks, number of bytes, and measured/computed data are output according to
the byte order specified with the BO command.

5.3 Output Format of BINARY Data

Advertising
This manual is related to the following products: