App index, Xi sets the a/d integral time, Xb sets the burnout detection – Yokogawa µR20000 User Manual

Page 64: Xj sets the rjc, Uc changes the dot color (dot model)

Advertising
background image

Commands

1

2

3

4

5

6

App

Index

IM 04P01B01-17E

4-21

p5 Hold/Not hold the relay

HOLD

Hold the relay output until an

alarm acknowledge operation is
executed

NONHOLD Reset the relay output when the

alarm is cleared.

p6 Hold/Not hold the alarm status display

HOLD Hold the display until an alarm

acknowledge operation is
executed

NONHOLD Clear the display output when

the alarm is cleared.

p7 Interval for the high limit on rate-of-change

alarm (01 to 15)

p8 Interval for the low limit on rate-of-change

alarm (01 to 15)

p9 Alarm hysteresis on measurement channels

(OFF, 0.1%, 0.2%, 0.3%, 0.4%, 0.5%, 0.6%,
0.7%, 0.8%, 0.9%, or 1.0%)

p10 Alarm hysteresis on computation channels

(OFF, 0.1%, 0.2%, 0.3%, 0.4%, 0.5%, 0.6%,
0.7%, 0.8%, 0.9%, or 1.0%)

Query

XA?

Example Enable fault diagnosis output. Disable reflash

and AND operations. Set the relay operation
to energize and hold. Set the alarm status
display to hold. Set the interval for the high limit
of rate-of-change alarm to 10, the interval for
the low limit of rate-of-change alarm to 12, the
measurement alarm hysteresis to 0.5%, and the
computation alarm hysteresis to OFF.
XAON,OFF,NONE,ENERGIZE,HOLD,HOLD,
10,12,0.5%,OFF

Description • The interval for the high limit of rate-of-change

alarm is equal to “scan interval × p7” and
the interval for the low limit of rate-of-change
alarm is equal to “scan interval × p8.”

• Parameter p10 can be specified on models

with the computation function (/M1 option).

• For the details on the settings, see the

µR10000 Recorder User’s Manual or the
µR20000 Recorder User’s Manual.

XI

Sets the A/D integral time.

Syntax

XI p1<terminator>
p1 Integration time of the A/D converter

AUTO

Automatically set in synch with
the power supply frequency.

50HZ

20 ms

60HZ

16.7 ms

100MS 100 ms (dot model)

Query

XI?

Example Set the A/D integral time to 50 Hz.

XI50HZ

Description • 100 ms is available only on the dot model.

• If Auto is specified when using the 24-VDC

power supply on a recorder with the 24-VDC/
AC power supply (/P1 option), the integration
time is fixed to 20 ms (50 Hz).

XB

Sets the burnout detection.

Syntax

XB p1,p2<terminator>
p1 Channel number
p2 The procedure taken when a sensor burnout

is detected
OFF Disable the burnout detection.
UP

Set the recording off the scale to the
right when a burnout is detected.

DOWN Set the recording off the scale to the

left when a burnout is detected.

Query

XB[p1]?

Example Set the recording off the scale to the right when a

sensor burnout is detected on channel 01.
XB01,UP

XJ

Sets the RJC.

When using the internal compensation circuit

Syntax

XJ p1,p2<terminator>
p1 Channel number
p2 Reference junction compensation selection

(INTERNAL)

Query

XJ[p1]?

Example Use the internal compensation circuit on channel

01.
XJ01,INTERNAL

When using an external RJC

Syntax

XJ p1,p2,p3<terminator>
p1 Channel number
p2 Reference junction compensation selection

(EXTERNAL)

p3 Compensation voltage (–20000 to 20000)

[µV]

Query

XJ[p1]?

Example Set the reference junction compensation of

channel 02 to external and set the compensation
voltage to 0 µV.
XJ02,EXTERNAL,0

Description The unit of p3 is µV (microvolts).

UC

Changes the dot color (dot model).

Syntax

UC p1,p2<terminator>
p1 Channel number
p2 Dot color

PURPLE
RED
GREEN
BLUE
BROWN
BLACK

Query

UC[p1]?

4.5 Basic Setting Commands

Advertising
This manual is related to the following products: