AM8060 Precision Thermometer

User’s Guide

13/43

4.5.2.4

If the sensor type is SPRT, the adjustable

parameters are Rtp, a, b, c, a4 and b4. The details of

these parameter definitions can be found from the ITS-

90.

As an example, calibration agencies provide following

parameters after SPRT is calibrated:

Rtp: probe’s resistance at triple point of water.

a4 and b4:low temperature range parameters.

a, b and c:high temperature range parameters (b and c can

be “0” for some instances. (Please refer to ITS-90 for more

information.)

If the sensor type is IPRT, the adjustable parameters are R0,

A, B and C:

The details of the parameter definitions can be found from

the industrial platinum resistance thermometer calibration

procedure.

As an example, calibration agencies provide following

parameters after IPRT is calibrated:

R0: the probe’s resistance at 0°C.

A and B: the temperature coefficients at temperature range

above 0°C.