AM8060 Precision Thermometer

User’s Guide

19/43

* 100 Ohm CALIBRATION

Connect a 100 Ohm

Standard Resistor

and Press Enter

Menu= Exit Enter = Yes

-------------------------------------------

5.1.4 The instrument will show the phrase “Connect a 100 Ohm

Standard Resistor and Press Enter” to indicate that a 100Ω

must be connected to sensor1 (S1) terminals as Figure3.
Users can exit the calibration by pressing

MENU

key to go back

to the main menu (*CALIBRATION MENU*).

5.1.5 After the probe is correctly connected, press

key to start

the calibration. The instrument will display “Waiting…”

while the resistance is calibrated. When the process is

finished, the screen will display the result as seen below:

* 100 Ohm CALIBRATION

100.0238

Rt = 111.0000

OHM

Menu= Exit

<^v>=Adj

100R Rs =

OHM

5.1.6 To calibrate the instrument, users must adjust the value

(100.0238 as shown above) to match the standard resistor’s

resistance. For example, if the resistance of the standard

resistor is 100.0038 ohm, users can move the underline
cursor to the desired position using

and

keys, and