AM8060 Precision Thermometer

User’s Guide

20/43

users can adjust the value by using

and

keys. The

adjusted result will appear like the picture below:

* 100 Ohm CALIBRATION

100.0038

Rt = 100.0038

OHM

Menu= Exit

<^v>=Adj

100R Rs =

OHM

5.1.7 It is required to wait for 10 seconds for the data to be stable.

Each second is shown with a blinking dot at the top left hand

corner of the display. Once this is completed, the next step

involves saving the calibrated data. To save the data, press
the

MENU

key and the third line of data will blink to indicate

the option to save.

5.1.8 To save this calibration value, press and hold the

key and

then press the

key. This will store the calibration data to

the instrument, and the program will go back to the main

menu (*CALIBRATION MENU*). If users DO NOT want
to save the data, press the

MENU

key and bring the program

back to main menu ( *CALIBRATION MENU*).

5.2 25 Ω calibration

5.2.1

The process of

25

Ω calibration is similar

to t

hat of 100 Ω

calibrationn. First, turn off the power of the instrument and