AM8060 Precision Thermometer

User’s Guide

18/43

5.1 100 Ω calibration

5.1.1 Turn off the power, put the measurement/calibration switch

to the “CAL”side (right side). Please see Figure 3 for details.

5.1.2 After turning on the power,AM8060 shows the calibration

main menu. It can be seen that there are two calibrations

required: 100 Ohm CALIBRATION and 25 Ohm

CALIBRATION.
The following is an illustration of 100 Ohm CALIBRATION.

* CALIBRATION MENU *

> 100 Ohm CALIBRATION

25 Ohm CALIBRATION

Enter = Yes

-------------------------------------------

5.1.3 Use

or

key on the keypad to select the calibration. A

“>” sign will be displayed in front of the selected line. For

example, the above figure shows that 100 Ohm
CALIBRATION has been selected. Pressing the

key will

start the calibration program, and the following display will

appear: