AM8060 Precision Thermometer

User’s Guide

22/43

resistance is calculated, and the following message will

appear:

25.9977

Rt =25.0000

OHM

Menu= Exit Enter = Yes

25R Rs =

OHM

* 25 Ohm CALIBRATION

5.2.6 The value on the third line is the parameter to be adjusted and

this value must match the standard resistance of the 25Ω

resistor. For example, the above figure shows the current

value as 25.9977, but the standard resistor’s value is 24.9977.
This value can be adjusted by using

or

key to move

the “_” (underscore) to the desired, and using

or

key

to adjust the value. After the adjustment, the instrument will

display:

24.9977

Rt =24.9977

OHM

Menu= Exit Enter = Yes

25R Rs =

OHM

* 25 Ohm CALIBRATION

5.2.7 It is required to wait for 10 seconds to ensure the reading is

stable. Each second is shown with a blinking dot at the top

left hand corner of the display. The next step is to save the

calibrated data to the instrument. To start the data saving,