AM8060 Precision Thermometer

User’s Guide

17/43

5 Instrument Calibration And Accuracy

Adjustments

AM8060 is designed for temperature measurement by measuring

the probe resistance and converting it into a temperature.

Consequently, the resistance measurement is very critical, and the

instrument must be calibrated correctly. Every AM8060 is

calibrated before shipment.

AM8060 needs regular calibration performed by qualified

professionals. One year calibration interval is recommended.

Users can calibrate AM8060 using standard resistors with

calibrated values.

Warning: It is required to set correct R0 (Rtpw) values in Sensor 1

(S1) before calibrations. To calibrate AM8060 in the range from

10 Ω to 50 Ω, R0 (Rtpw) needs to be set to 25 Ω. To calibrate

AM8060 in the range from 50 Ω to 375 Ω R0 (Rtpw) needs to be

set to 100 Ω.

In case of AM8060 out of tolerance, its accuracy can be adjusted.

Only qualified agencies, labs or institutes (who must be equipped

with calibrated 25 Ω and 100 Ω standard resistors) should adjust

the accuracy of AM8060.

The calibration should be done in an environment with room

temperature of 22±2°C.

Instrument calibration procedure: (No special sequence

for 100

Ω

and 25

Ω

calibration)