4.3.9 Capacitor Measurement (

)

CAUTION

To avoid damage to instrument
• Turn off the power to the circuit under test before starting measurement in order to prevent

any excessive voltage from being applied to the instrument.

• Before starting measurement, be sure to discharge the capacitor under check.

1) Turn the function switch to the “

”position.

2) Plug the test leads into the input terminals.
3) Open the test lead and press the REL key in 10nF range to adjust the

capacitance to zero. (The display shows “ 0.00 ”.)

4) Connect the test leads to the circuit under test and then read the value

when it stabilizes.

Test leads

Red

Black

Note

The value (zero adjustment) remains displayed until power-off.

4.3.10 Frequency Measurement (Hz)

CAUTION

To avoid damage to instrument

Turn off the power to the circuit under test before starting measurement in order to prevent any

excessive voltage from being applied to the instrument.

1) Turn the function switch to the “

V” position.

2) Press the SELECT key to select the range of frequency. (The

display shows the unit of frequency.)

3) Plug the test leads into the input terminals.
4) Connect the test leads to the under test and then read the

value when it stabilizes.

Test leads

Red

Black

<Voltage>