Carry out calibration of ranges in accordance with Table 1.

1) Turn the function switch from the OFF position to the mV position while pressing the

SELECT and RANGE keys at the same time.

The display shows the “ CAL ” symbol then the “ PASS ” symbol.
2) Press the SELECT key. (The display shows the “ - ” symbol.)
3) Press the HOLD key twice. (The display shows the “ - - - ” symbol.)
4) Press the RANGE key. (The display shows the “ mV ” symbol.)
5) Connect the instrument to the calibrator with the test leads.
6) Set the calibrator to Input value as an input to the instrument.
7) Press the HOLD key.
8) Be sure to confirm that the function switch and input terminal are set to the desired range.

Carry out calibration of other ranges by repairing steps 6) and 7) with reference to Table 1.

9) To quit calibration, turn the function switch back to the OFF position.

Note

Calibration should be started after fixing a range with the RANGE key.

Table 1. Calibration Table

Range

Input value

Range

Input value

DC600mV

600mV

AC6V (RMS) *1

6V 60Hz

DC6V

6V

AC6V (MEAN) *2

6V 60Hz

DC60V

60V

DC600V

600V

10nF

10nF

DC1000V

1000V

100nF

100nF

DC600μA

600μA

1μF

1μF

DC6000μA

6000μA

10μF

10μF

DC60mA

60mA

100μF

100μF

DC440mA

400mA

1000μF *3

1000μF

DC6A

6A

DC10A

10A

*1 : Calibration for all ranges by RMS value detection
*2 : Calibration for all ranges by MEAN value detection. (KEW1052 only)
*3 : Press the HOLD key 20 sec later after applying an input. It takes about 8 sec(max) to get

a readings stable. (A buzzer sounds.) Further operations should not be done until readings
becomes stable.