KYORITSU 1052 User Manual

Page 26

Advertising
background image

25

4.3.15 MIN/MAX/AVG Function (KEW1052 only)
The minimum value (MIN), maximum value (MAX) and average value (AVG) during
measurement are shown. (The range is fixed.) The average value is shown by dividing the

integrated record data by the number of recording times.
Pressing this key starts recording and at the same time the display shows “MIN”, “MAX” and
“AVG” to release AUTO POWER OFF.

<Recording time>

The timer is activated to show the elapsed time from the start and simultaneously the

renewed time for MIN/MAX is also recorded.

The elapsed time is displayed as follows:

0 sec. to 99 min. and 59 sec.: steps of 1 sec.

100 min. or more: steps of 1 min.

Press the HOLD key to stop recording. (The display shows the “ D•H ” symbol.)

<To confirm the recording time>

For confirming the recording time, press the MIN/MAX key.

Subsequent pressing of this key repeats to display the present minimum value (MIN),

maximum value (MAX) and average value (AVG).

Press the HOLD key once again to restart recording.
To cancel the confirming mode, hold down the MAX / MIN key for one second.

(“MAX” “MIN” “AVG” symbol disappears.)

Note

• No influence is exerted on the recorded data even if the test leads are disconnected while the

recording is stopped.

• If overload is recorded, the MIN or MAX display changes to “ OL ” display, resulting in

incorrect average data.

• For widely varying signal measurement, set the appropriate range in which the MAX or

MIN does not change to “ OL ” display.

Advertising
This manual is related to the following products: